ID: spacetop_ses-XX_sub-XXXX (each ID should be unique, otherwise it will throw an error)
Gender : M/F
Date of Birth : date of experiment (click on dropdown menu, calendar default is the date of experiment)
Doctor: Your initials
Also documented in DBIC setup
Medoc PC: Medoc TSA setup
Turn TSA-II on before experiment, for safety test to pass
check green light on TSA-II
check connection between TSA-II & medoc-PC (USB!!)
Turn Medoc PC on and open Acknowledge software (admin/admin)
Click participant, test, and make sure “external control” is on.
* External control “on”: External control > check box
plug in participant information (use BIDS id)
Medoc PC - Stim PC communictaion
Check if all of the ethernet cables are connected [ #LINK configuration ]
Test 1: In the SPACETOP PC terminal, type ping 10.64.1.10 This is the ip address of the medoc PC. we’ll make sure that the spacetop can send bytes to medoc
Test 2: turn on medoc main station program. Go into external control mode. In the SPACETOP PC, load matlab ptb3-matlab. In the command window, main(‘10.64.1.10’, 20121, 1, 100) # select Program main(10.64.1.10’, 20121, 4, 100) # Trigger Program
run testmain.m, which will basically do Test 2 for you.